Insulation work of medical device shell

Release time:2023-03-30

Medical diagnosis, measurement and treatment equipment using AC power supply may cause leakage current due to improper grounding and electrical insulation, which may expose patients and even medical staff to the risk of electric shock, burns, internal organ damage and arrhythmia. So do a good jobMedical device shellInsulation work.

美容儀器殼9.jpg


1. The equipment shall be manufactured and sealed to prevent contact with live parts and parts that may be live in a single fault state. If the bulb can be replaced without tools, it must be ensured to prevent contact with the live parts of the lamp when loading and unloading the bulb.


2. The position or size of any hole on the top cover of the medical device shell shall ensure that the test rod with a diameter of 4 mm and a length of 100 mm will not touch live parts when it is freely and vertically suspended in the access hole throughout its length.


3. When the handle, knob, control lever, etc. are removed, the conductor part of the operating mechanism of the control device can be touched. When the open circuit test voltage is not greater than 50V AC and the test current is not less than 1a, the resistance measured to the protective grounding terminal should not be greater than 0.2 Ω; Or it should be isolated from live parts by adopting a method of 17g in GB 9706.1-2007.


4. If the components with line voltage above 25V AC or 60V DC in the shell of the medical device cannot be disconnected from the power supply through the external main switch or the plug device that can be pulled out at any time, additional covers shall be used for protection, so that contact can be prevented even after the shell is opened, or when the spaces are separated and arranged, the "live" mark shall be clearly marked.


5. In order to prevent contact with live parts, the shell must be removed only with tools, otherwise an automatic device should be used when opening or removing the shell, so that these parts are not used.


6. When the device has a preset controller, it can be adjusted by the user with tools during normal use. The hole opened for adjusting these control devices shall be designed so that the adjusting tool cannot contact the basic insulation or any live parts in the hole, or any parts isolated from the grid power supply part only through the basic insulation but without grounding protection.


Copyright: Shenzhen Zhongyuan Plastic Co., Ltd. | mold processing customized manufacturer. If you want to know the processing price of the factory, how much is it, and which is good, please contact us Yue ICP Bei No. 17060480
主站蜘蛛池模板: 日本在线观看 | 亚洲国产一区二区在线 | 色伊人国产高清在线 | 四虎新地址 | 在线亚洲精品自拍 | 伊人五月婷婷 | 日本韩国伦理片 | 青春草视频在线免费观看 | 五月亭亭六月丁香 | 欧美综合影院 | 亚洲欧美男人天堂 | 西西gogo午夜高清大胆 | 欧美一级片观看 | 日韩欧美在线观看综合网另类 | 天天干天天射天天 | 西西人体大胆77777视频 | 三级www | 日韩成人免费aa在线看 | 综合网 色天使 | 天天做日日干 | 午夜激情在线观看 | 日韩国产在线 | 欧美在线免费观看视频 | 亚洲福利国产 | 日本成人福利视频 | 日本一区二区三区免费在线观看 | 偷自视频视频区免费 | 最新午夜 | 午夜久久久久久亚洲国产精品 | 亚欧美| 天天干天天插天天 | 欧美香蕉视频在线观看 | 日本成人免费在线观看 | 青娱乐手机在线视频 | 欧洲a级片 | 四虎永久网址影院 | 日韩手机专区 | 亚洲自拍小视频 | 青青草免费观看视频 | 亚洲欧美在线免费观看 | 日日夜夜精品免费视频 |